8-1
8 PARAMETERS SETTING MODE
8 PARAMETERS SETTING MODE
8.1 Parameter Setting Options
8.1.1 Parameters for Measurement and Display
Menu Selecting Item Contents
Ang. Unit deg / gon / mil
Select degree(360°), gon(400G) or mil (6400M) for the
measuring angle unit to be shown on the display.
Min.Angle OFF / ON
Select the minimum display angle reading ON or OFF.
GTS-821A [ OFF:1" / ON:0.5" ]
GTS-822A [ OFF:1" / ON:0.5" ]
GTS-823A [ OFF:5" / ON:1" ]
GTS-825A [ OFF:5" / ON:1" ]
Tilt OFF / 1axis / 2axes
Select the tilt sensor option for OFF, (1axis) vertical only or
(2axes) vertical and horizontal.
Err. corr. OFF / ON
Select the error correction ON or OFF for collimation and error
adjustment.
Note
:
Perform this item after complete section 9.4.
For more information, refer to Section 9.4 Adjustment of
Compensation Systematic Error of Instrument” and Section
9.5 “Showing Constant List and Switch ON/OFF
Compensation Systematic Error of Instrument” .
V -0 Zenith / Level Select the vertical angle reading for Zenith 0 or Horizontal 0.
HAmem OFF / MEM. ON
It is able to retain presetting angle after turning power
off.(MEMORY)
Note
:
After changing this parameter, turn the power switch off once.
Turn
Fine / Normal /
Coarse
Select the stopping accuracy of automatic rotation to required
angle.
Fine:3" Normal :5" Coarse:10"
Auto.Aim
Fine / Normal /
Coarse
Select the accuracy of automatic collimation.
Prog
Std
Mem Comm
Adj
Para
2003-10-10 15:30:40
PARAMETERS SETTING MODE
[Press [F6] key.]
In this mode, setting of parameters regard with measuring, displaying and
communications will be done.
When a parameter is changed and set, the new value is stored into memory.
Press [F6] key from the main menu icons, the following display will be shown.
The parameter modes is classified in Measurement and Communication.
Parameters
F1 Measurement
F2 Communication
F3 Password
Vista de página 91